Suppression of Coriolis error in weak equivalence principle test using 85Rb–87Rb dual-species atom interferometer
Duan Wei-Tao1, 2, He Chuan1, 2, Yan Si-Tong1, 3, Ji Yu-Hang1, Zhou Lin1, 2, 3, †, Chen Xi1, 2, 3, Wang Jin1, 2, 3, ‡, Zhan Ming-Sheng1, 2, 3
State Key Laboratory of Magnetic Resonance and Atomic and Molecular Physics, Innovation Academy for Precision Measurement Science and Technology, Chinese Academy of Sciences–Wuhan National Laboratory for Optoelectronics, Wuhan 430071, China
School of Physics, University of Chinese Academy of Sciences, Beijing 100049, China
Center for Cold Atoms Physics, Chinese Academy of Sciences, Wuhan 430071, China

 

† Corresponding author. E-mail: lzhou@wipm.ac.cn wangjin@wipm.ac.cn

Project supported by the National Key Research and Development Program of China (Grant No. 2016YFA0302002), the National Natural Science Foundation of China (Grant Nos. 91736311 and 11574354), Strategic Priority Research Program of the Chinese Academy of Sciences (Grant No. XDB21010100), and the Youth Innovation Promotion Association of the Chinese Academy of Sciences (Grant No. 2016300).

Abstract

Coriolis effect is an important error source in the weak equivalence principle (WEP) test using atom interferometer. In this paper, the problem of Coriolis error in WEP test is studied theoretically and experimentally. In theoretical simulation, the Coriolis effect is analyzed by establishing an error model. The measurement errors of Eötvös coefficient (η) in WEP test related to experimental parameters, such as horizontal-velocity difference and horizontal-position difference of atomic clouds, horizontal-position difference of detectors, and rotation compensation of Raman laser’s mirror are calculated. In experimental investigation, the position difference between 85Rb and 87Rb atomic clouds is reduced to 0.1 mm by optimizing the experimental parameters, an alternating detection method is used to suppress the error caused by detection position difference, thus the Coriolis error related to the atomic clouds and detectors is reduced to 1.1× 10−9. This Coriolis error is further corrected by compensating the rotation of Raman laser’s mirror, and the total uncertainty of η measurement related to the Coriolis effect is reduced as δη = 4.4 × 10−11.

1. Introduction

Atom interferometer (AI)[1] is a new type of precision measuring instrument, which has been widely used in precision measurement of gravity,[211] gravity gradient,[8,1214] rotation,[1519] and test of weak equivalent principle (WEP).[2029] Coriolis effect is an important error source in gravity measurement experiments using AIs.[11,3034] To compensate the effect of the Coriolis effect in atom interferometry, several effective methods have been used in single-species AI-based gravity measurements. Peters et al.[33] used the interference signal itself to find the correct position of the detection region (where the Coriolis effect vanishes) by introducing additional, well-defined rotations. Louchet-Chauvet et al.[34] corrected Coriolis shift by carrying out measurements with opposite orientations with respect to the Earth’s rotation vector direction. Lan et al.[35] used a piezoelectric tube (PZT)-controlled tip-tilt mirror to regulate the orientation of the Bragg beam and compensate for the Earth’s rotation. Different from the configuration of the single-species AI, the dual-species AI used for WEP tests can achieve time- or space-synchronized interference measurement using different species atoms, which has good performance in common-mode-noise suppression. However, due to the differences in the initial positions and velocities of the dual-species atomic clouds, the Coriolis effect related to the horizontal motion of the atomic clouds is still a major error source in AI-based WEP tests.[21,22] In 85Rb–87Rb dual-species AI, we achieved common-mode-noise suppression by the four-wave double-diffraction Raman transition (4WDR) scheme,[24] which greatly reduced contributions from laser phase noise, ac Stark shift, and quadratic Zeeman shift. But, the Coriolis effect error, which was introduced by the small inconsistent of the centroid positions and velocities of 85Rb and 87Rb atomic clouds, was not effectively suppressed, the uncertainty of this error was as high as 2.9 × 10−8,[24] which was the largest error term in the experiment at that time. To effectively improve the precision of WEP test using the 4WDR dual-species AI, we study the suppression of the Coriolis error in this paper. We first theoretically analyze the phase shift factors related to the Coriolis effect in the 4WDR AI, then calculate the Coriolis errors related to the experimental parameters, such as the velocity differences, position differences of atomic clouds, the position differences of detectors, and the rotation rate of Raman laser’s mirror. Based on the analysis, we find experimental means to suppress the Coriolis effect. By optimizing the background magnetic field, the parameters of the cooling laser beams, and the loading time of the magneto-optical trap (MOT), the position difference between 85Rb and 87Rb atomic clouds is reduced. By using the alternating detection method (ADM), the uncertainty caused by the position fluctuation of the dual-species atomic clouds is suppressed. By two-dimensional (2D) rotation compensation for Raman laser’s mirror, the total uncertainty of η measurement related to the Coriolis effect is reduced to δη = 4.4 × 10−11.

2. Theoretical model

To quantitatively investigate the Coriolis error in the 4WDR AI and provide guidance for experiments, we establish an error model to analyze the dependence of the Coriolis phase shift on the velocity, position of the atomic clouds, and the position of the detectors (the velocity and position discussed in this paper are in the horizontal direction), then theoretically simulate the performance of rotation compensation of Raman laser’s mirror.

2.1. Error model

The Eötvös coefficient in 85Rb–87Rb dual-species AI-based WEP test is[24]

where gi (i = 87, 85) is the gravitational acceleration of 87Rb or 85Rb atoms. The phase shift is simply described as

where (i = 87, 85) is the effective wave vector of Raman laser for 87Rb or 85Rb atoms, and T the interval between the Raman pulses. According to Eqs. (1) and (2), the relation between η and phase shifts is

Usually, the value of η measured by experiments consists of three parts, η = η0 + Δη ± δη, where η0 is the ideal value of the Eötvös coefficient, Δη the systematic error, and δη the uncertainty of Δη.

The schematic diagram of a 85Rb–87Rb dual-species AI is shown in Fig. 1. Figure 1(a) shows the schematic diagram of relevant hyperfine-levels of rubidium atoms. Take the 85Rb atom as an example to briefly explain the atom interference process. As shown in Fig. 1(b), the atoms in initial state |F = 2〉 interact with π/2–ππ/2 Raman pulses to implement interference loop. The wave vectors of the three Raman lasers are k1, k2, and k3. The atom interference loop is spatially symmetric because Raman pairs (k1, k3) and (k2, k3) supply recoil momenta in opposite directions. The first π/2-pulse is for transferring atoms from state 85Rb|F = 2, 0〉 to states 85Rb|F = 3, ℏkeff〉 and 85Rb|F = 3,–ℏkeff〉 at site A, where ℏkeff and –ℏkeff are the recoil momenta of atoms obtained from the Raman pulse. The π-pulse is for reflecting atoms in different monument states 85Rb|F = 3, ℏkeff〉 and 85Rb|F = 3, –ℏkeff〉 at sites B and C, where the populations of different monument states exchange. The second π/2-pulse is for combing atoms in states 85Rb|F = 3, ℏkeff〉 and 85Rb|F = 3, –ℏkeff〉 at site D. The detected population of 85Rb| F = 2〉 or 85Rb| F = 3〉 at D depends on the phase shifts of the Raman pulses. Similarly, for 87Rb atoms, the initial state is 87Rb|F = 1, 0〉 and the interference states are 87Rb| F = 2, ℏkeff〉 and 87Rb| F = 2, –ℏkeff〉. The wave vectors of the three Raman lasers are k1, k2, and k4.

Fig. 1. Schematic diagram of 85Rb–87Rb dual-species atom interferometer. (a) Related sub-levels of rubidium atoms. Raman lasers with frequencies ω1, ω2, and ω3 are used for 85Rb atoms, while those with frequencies ω1, ω2, and ω4 are for 87Rb atoms; ω1 and ω2 are detuned to the blue side of transitions 85Rb|F = 3〉 to |F′ = 4〉 and 87Rb|F = 2〉 to |F〉 = 3〉. (b) Schematic diagram of 4WDR interference path for 85Rb atoms, k1, k2, and k3 are wave vectors of Raman beams ω1, ω2, and ω3, respectively. For 87Rb atoms, states 87Rb|F = 1〉 and 87Rb|F = 2〉 are involved, and k3 is replaced by k4.

In the local laboratory (Wuhan, 30.54°N) coordinate system, assume that the positive direction of the z-axis is upward along the plumb line. The y-axis is the direction of the unit vector in the horizontal plane from south to north, x-axis is the direction of the unit vector in the horizontal plane from west to east. Considering the Earth as a regular sphere, the rotation angular vector of the Earth is written as

The projection of the Earth’s rotation rate in the laboratory’s y-axis is Ωy. We define Ωy = Ω0 as the reference point for rotation compensation, Ω0 = 62.8 μrad/s. The trajectories of atoms are expanded as time-varying polynomials

where, ri (i = 0, 1, 2, …) are time-independent parameters. The dynamic equation of the atom is[36]

where g is the local gravitational acceleration, R the connecting line between the Earth’s center point and the coordinate origin of the laboratory system, and Γ the local gravity gradient tensor. The total Coriolis phase shift of the interferometer is

where ν0 is the average horizontal-velocity of the atoms.

In the experiment, the average velocity of the atomic clouds measured by a detector is related to the detection position.[33,37] To quantitatively describe the velocity distribution of atoms in the detection area, we directly calculate the probability of atoms in the detection area. At the initial moment, the state of atom can be described by the product of the initial velocity distribution and the initial position distribution. Assume the atoms are inside the detection area at time t, the convolution (that is the equivalent velocity distribution of the atoms in the detection area) of atom’s position and velocity can be obtained by full-space integration. Thus, we calculate the Coriolis phase shift in differential-measurement of gravity using dual-species AI under various atomic parameters and detection parameters.

The Coriolis effect was compensated by changing the orientation of Raman laser’s wave-vector in single-species AI.[35] Here, we apply this rotation compensation method to the dual-species AI. Consider that the incident Raman beam is reflected by a PZT-driven mirror. Let ki be the wave-vector of the incident beam, ko the wave-vector of the exit beam, and n the unit normal vector of the mirror, then ko = ki – 2(kin)n.[38] In the mirror coordinate system, the west–east diameter in the mirror surface is defined as x′-axis, and the diameter orthogonal to both x′-axis and n is y′-axis. When the 2D compensation angle (θx′,θy′) is applied to each Raman pulse, the compensated phase shift is

where r(0), r(T), and r(2T) are the atomic positions at time 0, T, and 2T, respectively; θx′ and θy′ are the rotation angles of the PZT around x′ – and y′-axis, respectively.

In the laboratory coordinate system, the actual rotation compensation performance can be evaluated by the contrast of interference fringes,[35] that is, when the fringe contrast reaches the maximum value, the corresponding mirror rotation rate just compensates the Earth’s rotation effect, the relevant phase shift is represented as

where ci (i = 1, 2, 3, 4) is the coefficient of parameter i; υx and υy are the velocities of atoms in the x and y directions, respectively; x and y are the positions of atoms in the x and y directions, respectively. The fringe contrast is expressed as

where σj (j = υx, υy, x0, y0) is the standard deviation of Gaussian distribution of j. In the following, we consider the actual experimental conditions and analyze the Coriolis error in 85Rb–87Rb dual-species AI-based WEP test.

2.2. Parameters of atomic clouds and Coriolis error

Due to the differences in mass and hyperfine levels, the average velocity, center position, size, and temperature of 85Rb and 87Rb atomic clouds are also different. Suppose the atoms are launched 2 m up and fall to the detection area after 1.2 s, the area of the detection is 10 mm × 10 mm, two detectors for 85Rb and 87Rb atoms are 30 mm apart, and the atomic clouds are symmetrically distributed in the vertical direction. We calculate the dependence of the Coriolis shift on the difference of the initial velocity and position between 85Rb and 87Rb atomic clouds.

The simulated error of η versus atomic cloud parameters is shown in Fig. 2. Figure 2(a) shows the dependence of Δη on the velocity difference between 85Rb and 87Rb atomic clouds, where the temperatures of the atomic clouds are selected as 10 nK, 100 nK, and 3 μK, and the sizes of the atomic clouds are selected as 0.1 mm, 1 mm, 3 mm, and 5 mm. Figure 2(b) shows the dependence of Δ η on the position difference between 85Rb and 87Rb atomic clouds under various sizes and temperatures. It can be seen that with the increase of the temperature and size of the atomic clouds, Δη varies differently with the initial velocity difference and position difference. For ultra-cold (10 nK, 0.1 mm) 85Rb and 87Rb atomic clouds, the Coriolis effect caused by the initial velocity difference is much larger than that of cold (3 μK, 3 mm) atomic clouds, while the Coriolis effect caused by the position difference is smaller than that of cold (3 μK, 3 mm) atomic clouds. This is because when the atoms freely diffuse for 1.2 s, the sizes of the 10 nK clouds are small, and the average velocity difference between the two atomic clouds at the detection area does not change much from that at launching point, while the 3 μK clouds have large sizes which are selectively detected by the position-fixed detectors, only 85Rb and 87Rb atoms with smaller velocity difference synchronously enter the detection area. Therefore, the Δη value simulated by 3 μK 85Rb and 87Rb atoms varies slowly with the initial velocity difference. However, when the initial positions are different, the average horizontal-velocities of 3 μK 85Rb and 87Rb atomic clouds which enter the same detection area are different, and their difference is much larger than that of the 10 nK clouds, the error of η value simulated by 3 μK 85Rb and 87Rb atoms varies quickly with the initial position difference. For typical 85Rb and 87Rb atomic clouds with size of 3 mm, temperature of 3 μK, position difference of 0.1 mm, and velocity difference of 1 mm/s, the systematic error of the Coriolis effect is 1.1 × 10−9 and that caused by the velocity difference is –5.8× 10−10.

Fig. 2. Dependence of Δ η on the initial parameters of dual-species atomic clouds. (a) Δη versus initial velocity difference between 85Rb and 87Rb atomic clouds. (b) Δη versus initial position difference between 85Rb and 87Rb atomic clouds. The dot line is the data for 85Rb and 87Rb atomic clouds with size of 0.1 mm and temperature of 10 nK, the dash line is for 1 mm, 0.1 μK atomic clouds, the dash-dot line is for 3 mm, 3 μK atomic clouds, the solid line is for 3 mm, 3 μK 85Rb atomic cloud and 5 mm, 3 μK 87Rb atomic cloud.

In practical experiment, when the size of the detection area is fixed, for ultra-cold 85Rb and 87Rb atomic clouds with an initial temperature of 10 nK, it is necessary to accurately control and reduce their velocity difference to suppress the Coriolis error. For cold atomic clouds with an initial temperature of several microkelvin, it is necessary to precisely control and reduce their position difference to eliminate the Coriolis error.

2.3. Detector location and Coriolis error

Since the Coriolis effect is proportional to the horizontal-velocity of the atomic clouds, different detecting position leads to a difference in horizontal-velocity selection. This detection effect[37] introduces a Coriolis error. Assuming 85Rb and 87Rb atomic clouds are detected simultaneously by two different detectors, we simulate the dependence of Δη on the position difference of detection, the results are shown in Fig. 3. When the initial position and velocity of the atomic clouds are determined (corresponding to the case where the temperature is 3 μK and the size of the detection area is 10 mm × 10 mm), if the detection positions of 85Rb and 87Rb atoms differ by 1 mm, then the measurement error of η value is about 2 × 10−9. If we exchange the two detectors for 85Rb and 87Rb atoms and average the data collected before and after exchanging, the average value is no longer sensitive to the detection position, the measurement error of η value is ignorable as shown by the solid line in Fig. 3. This means that under the ideal condition where the initial positions of the atomic clouds completely overlap, the ADM can effectively suppress the error caused by the detection position.

Fig. 3. Error of η versus the difference of detection position. The dash and dot lines are cases before and after alternating detectors, the solid line is the average data for both.
2.4. Suppressing Coriolis error by rotation compensation

To effectively correct the Coriolis errors introduced by the position fluctuation of atomic clouds, the initial position difference and initial velocity difference of dual-species atomic clouds, it is necessary to compensate the tilt angle of Raman laser’s mirror caused by the Earth’s rotation. Let Ωm represent the actual rotation rate of Raman laser’s wave vector around y-axis, the theoretically simulated relation between Ωm and the fringe contrast is shown in Fig. 4, where all atoms are assumed to participate in the interference process and contribute to the signal. It shows that the rotation compensation of Raman laser’s mirror corrects the total Coriolis error and significantly improves the fringe contrast. When Ωm = –Ω0, the fringe contrasts of atoms with different temperatures reach the maximum value. When Ωm ≠ –Ω0, the fringe contrast decreases rapidly with increasing atom temperature. For atoms with temperature of 3 μK, if no rotation compensation is performed (Ωm = 0), the fringe contrast is close to 0. For atoms with temperature of 0.3 μK, the fringe contrast is still higher than 60% even without rotation compensation.

Fig. 4. Fringe contrast (T = 203 ms) versus mirror’s rotation rate. The influence of the detector was not considered here. The dash-dot line is for atoms with temperature of 10 nK, the rotation compensation is not obvious. The dot line is for 0.3 μK atoms, the rotation compensation is effective in the range of –4Ω0 to + 2Ω0, the solid line is for 3 μK atoms, the rotation compensation is effective in the range of –1.5Ω0 to + 0.5Ω0.

The simulated dependence of Δη on mirror’s rotation rate is shown in Fig. 5, where both the position difference of atomic clouds and the inconsistencies of detectors are considered, the temperature of the atoms is 3 μK, the size of the atomic clouds is 3 mm, and their horizontal center positions deviate by 0.1 mm. According to Fig. 5, when Ωm = –Ω0, the curves obtained by two detection methods intersect, and the ordinate of the crossing point is 0. That is, at the ideal value of rotation compensation, the fringe contrast is the best, the error of η measurement is the smallest. From Fig. 5 we find that the errors caused by both detection methods (dot line and dash lines) and the mismatch of atomic clouds (slid line) have linear relationship with the mirror rotation angle, therefore, the optimal rotation rate can be determined by the fringe contrasts.

Fig. 5. Error of η-value versus rotation rate of the Raman laser’s mirror. Dash and dot lines are data for alternating detectors. Solid line is the average data of two detection methods, which is the contribution of position difference of 85Rb and 87Rb atomic clouds.

We also simulate the rotation compensation around y-axis for Raman laser’s mirror. The dependence of fringe contrast on the interval of Raman pulses (T) is shown in Fig. 6. If T = 200 ms, the fringe contrast with rotation compensation is increased from 67% to 99%; when T = 1 s, the fringe contrast with rotation compensation is increased from 0 to 71%. The rotation compensation is necessary for situations with longer Raman pulse intervals in long baseline AI.

Fig. 6. Fringe contrast versus the interval of Raman pulses (T). The dash line is the simulation without rotation compensation, and the solid line is the simulation with rotation compensation rate at 0.9Ω0.
3. Experimental measurement

The above simulations have three implications: (1) controlling and reducing the initial position and initial velocity differences of the atomic clouds is a necessary means to reduce the Coriolis error; (2) the ADM is useful to reduce the measurement error introduced by detection effect; (3) rotation compensation of Raman laser’s mirror is necessary for correcting the Coriolis error. The experimental setup and experimental results are described in detail as below.

3.1. Experimental setup

In our previous experiments,[24,39] due to the difference between the initial velocities of 85Rb and 87Rb atomic clouds, the resulting Coriolis uncertainty was δη = 2.9 × 10−8.[24] To improve the positional coincidence of atomic clouds, reduce detection error, and effectively compensate the Coriolis effect, here we add a 2D-PZT turntable (PI S-330.2SH) to control Raman laser’s mirror. The experimental setup is shown in Fig. 7. The Raman laser’s mirror is mounted on the 2D-PZT turntable with a cross angle of 45° to the horizontal plane. The 2D-PZT turntable can perform rotation modulation on x′ - and y′-axis, it has a rotation resolution of 20 nrad and an angle tuning range of 20 mrad. Four Raman laser beams (with frequencies of ω1, ω2, ω3, and ω4) are divided into two groups. One group (ω2, ω3, and ω4) is reflected by the mirror and propagates upward from the bottom window of the MOT chamber. Another group (ω1, ω3, and ω4) propagates downward from the top window of the vacuum chamber. The center of the detection area is located 25.4 cm above the center of the MOT chamber. Detectors A and B are installed on the north and south sides of the detection area, respectively. They are 30 mm apart in the vertical direction.

Fig. 7. Experimental setup. The Raman laser beams for 87Rb atoms (ω1, ω3, and ω4) propagate from the top window of vacuum chamber to the bottom and those for 85Rb atoms (ω2, ω3, and ω4) propagate from the bottom window of vacuum chamber to the top. PZT: piezoelectric tube. MOT: magneto-optical trap.

The MOT simultaneously traps and launches up 85Rb and 87Rb atoms to form dual-species atom fountains. The height of the fountains is 2 m. During the launching process, the atoms undergo velocity-selection and state-preparation, they then enter the magnetically shielded region, where the 4WDR scheme is used to implement a dual-species AI. After the interference process, the 85Rb and 87Rb atoms fall into the detection area and are detected by two detectors. The duration of the first Raman pulse is 31 μs, the interval of the Raman pulses is adjusted to 203.164 ms, so that the offset phase difference between the dual-species interference fringes is 20.5π, this is helpful for fitting the actual phase difference between 85Rb and 87Rb atoms by the ellipse fitting method.[40]

To improve the coincidence of interference trajectories of the two species atoms, the background magnetic field, the frequency stability, and the intensity stability of the lasers are optimized. This makes the atomic cloud in the MOT closer to a sphere. In the experiment, the size of the 87Rb atomic cloud increases with MOT loading time t. When t = 2 s, the sizes of 87Rb and 85Rb atomic clouds are similar. When t > 5 s, the size of 87Rb atomic cloud is twice that of 85Rb. Therefore, the loading time of the atoms in the MOT is adjusted to 2 s, so that the position difference of the two species of atomic clouds is reduced to the order of 0.1 mm. Atoms undergo vertical-velocity-selection before they enter the interference area, the control accuracy of vertical-velocity is 0.2 mm/s. These means effectively reduce the influence of the Coriolis effect, wave-front distortion, gravity gradient, and other factors.

In the experiment, the detection error comes from several factors, such as the size and position of detection beam, the angle of fluorescence collection, and the installation accuracy of detectors. The installation accuracy of the two detectors in west–east direction is better than 1 mm, and that in south–north direction is greater than 1 mm. The angle of the detectors in south–north direction is evaluated and adjusted according to the fluorescence signal. Affected by the fluctuations of the magnetic field, laser frequency, and laser intensity, the positions of the atomic clouds fluctuate. To solve these problems, a real-time ADM is used. In the first detection cycle, detector A collects the fluorescence signal of 85Rb atoms, detector B collects the signal of 87Rb atoms; in the second detection cycle, the two detection beams are exchanged, and detector A collects the signal of 87Rb atoms, detector B collects that of 85Rb atoms. These two detection steps are performed cyclically, which effectively suppresses the error caused by detection and atomic clouds’ position.

The rotation compensation of Raman laser’s mirror is achieved by the 2D-PZT turntable. When π/2, π, and π/2 Raman pulses reach the mirror, their wave vectors are rotated 0, ΩmT, and 2ΩmT by adjusting the tilt angle of the mirror, respectively. Because the normal direction of the mirror is orthogonal to the x′- and y′-axis of the PZT, it has a cross angle of 45° to both y-axis (south–north direction) and z-axis (vertical upward). The minor angle error between the project of the y′-axis of the PZT and the y-axis affects the measurement accuracy. The laboratory is located 16 m underground, a laser gyroscope is used to calibrate the south–north direction with a positioning accuracy of 2°, the deviation between the projection of y′-axis in the horizontal plane and the y-axis is within 5° (limited by the error of the PZT installation and positioning).

To quantitatively calibrate the error caused by the mirror, rotation compensation is applied for the mirror around x-axis and y-axis. First, the rotation around y-axis is compensated to obtain the optimal compensation angle, then the rotation around x-axis is compensated. The compensation angle in x-axis is very small, but it causes excessive compensation in y-axis. Finally, the total compensated error is evaluated by considering the error in x- and y-axis, and the theoretical estimation of overcompensation in y-axis.

3.2. Experimental results and discussion

After 2D rotation compensation is performed on Raman laser’s mirror, the fringe contrast is significantly improved. The experimental results of fringe contrast versus mirror’s rotation rate are shown in Fig. 8, where two sets of data by exchanging detectors A and B are included, and each set of data includes fringe contrasts of both 85Rb and 87Rb atoms. The rotation rate around y-axis varies from –2.5 Ω0 to 0.5Ω0, and that around x-axis varies from –1.5 Ω0 to 1.5 Ω0. The best compensation value and standard deviation of rotation rate around y-axis is (–1.036 ± 0.025)Ω0, and that around x-axis is (0.068 ± 0.021)Ω0. It should be noted that, for atomic clouds with initial temperature of 3 μK, only atoms with a relatively lower horizontal-velocity were detected by the detectors, and the actual measurements correspond to an atomic temperature of 100 nK level.

Fig. 8. Atom interference fringes contrasts versus mirror’s rotation rate. (a) Rotation compensated around y-axis, 85Rb atoms detected by detector A and 87Rb atoms detected detector B. (b) Rotation compensated around y-axis, 85Rb atoms detected by detector B and 87Rb atoms detected detector A. (c) Rotation compensated around x-axis, 85Rb atoms detected by detector A and 87Rb atoms detected detector B. (d) Rotation compensated around x-axis, 85Rb atoms detected by detector B and 87Rb atoms detected detector A.

The experimental results of Δη versus the rotation rate of Raman laser’s mirror are shown in Fig. 9, where the rotation compensation coefficients around y-axis and x-axis are –1.1 × 10−9/Ω0 and –1.6 × 10−9/Ω0, respectively. The coincidences of the 85Rb and 87Rb atomic clouds in south–north and west–east directions are similar. However, the error caused by the detection difference in west–east direction is smaller than that of the position difference of 85Rb and 87Rb atomic clouds, while in south–north direction it is 3.8 times of that of the position difference of 85Rb and 87Rb atomic clouds. The intersection of two sets of data by exchanging detectors A and B according to x-axis is near 0 (Fig. 9(b)), while that according to y-axis is far from 0 (Fig. 9(a)). This is because there is still a small installation error in the PZT rotation compensation system. When the mirror is rotated around x-axis, it will also cause a small rotation component in y-axis, so that the intersection of the two sets of data according to y-axis has an offset from the reference point Ωm = –Ω0. From the fringe contrast data compensated around the x-axis, the difference angle between y′-axis of the PZT and y-axis is calculated to be –2.3°, which is less than the installation accuracy (5°). Due to the existence of this difference angle, a correction of –1.1 × 10−11 is added to the data after rotation compensation around the y-axis.

Fig. 9. Error of η value versus rotation rate of Raman laser’s mirror. (a) Rotation compensation around y-axis. (b) Rotation compensation around x-axis. When compensating for rotation around x-axis, the compensation value around y-axis is fixed at Ωm = –Ω0. The coefficient of rotation compensation around y-axis is –1.1 × 10−9/Ω0 and that around x-axis is –1.6 × 10−9/Ω0.

Based on the above data, the uncertainty of rotation rate around y-axis is 0.025Ω0 and that of Coriolis error is 1.1 × 10−9/Ω0 × 0.025 Ω0 = 2.8 × 10−11; the uncertainty of rotation rate around x-axis is 0.021 Ω0 and that of Coriolis error is 1.6 × 10−9/Ω0 × 0.021 Ω0 = 3.4 × 10−11. The total uncertainty of Eötvös coefficient is δη = 4.4 × 10−11.

4. Conclusion and perspectives

In summary, the Coriolis error in AI-based WEP test was analyzed theoretically and measured experimentally. An error model suitable for Coriolis effect in the 4WDR dual species AI was built, the contribution of key experimental parameters to the Coriolis phase shift was analyzed. The Coriolis phase shift introduced by position mismatch of atomic clouds and position difference of detectors was suppressed by the ADM, and the Coriolis error caused by the inconsistent initial velocity of the dual-species atomic clouds was corrected by compensating the rotation of Raman laser’s mirror, and the uncertainty of η measurement in 85Rb–87Rb dual-species AI-based WEP test was improved to δη = 4.4 × 10−11. The experimentally measured error terms are consistent with the theoretical simulations. In addition, the dependence of fringe contrast on 2D rotation compensation was simulated under the condition of free evolution time of 1 s, which will be of significance for long-baseline AI-based higher-precision WEP tests in future.

Reference
[1] Wang J 2015 Chin. Phys. 24 053702
[2] Kasevich M A Chu S 1991 Phys. Rev. Lett. 67 181
[3] Peters A Chung K Y Chu S 1999 Nature 400 849
[4] Zhou L Xiong Z Y Yang W Tang B Peng W C Wang Y B Xu P Wang J Zhan M S 2011 Chin. Phys. Lett. 28 013701
[5] Merlet S Bodart Q Malossi N Landragin A Pereira Dos Santos F Gitlein O Timmen L 2010 Metrologia 47 L9
[6] Poli N Wang F Y Tarallo M G Alberti A Prevedelli M Tino G M 2011 Phys. Rev. Lett. 106 038501
[7] Hu Z K Sun B L Duan X C Zhou M K Chen L L Zhan S Zhang Q Z Luo J 2013 Phys. Rev. 88 043610
[8] Bidel Y Carraz O Charrière R Cadoret M Zahzam N Bresson A 2013 Appl. Phys. Lett. 102 144107
[9] Wu B Wang Z Cheng B Wang Q Xu A Lin Q 2014 Metrologia 51 452
[10] Wang S K Zhao Y Zhuang W Li T C Wu S Q Feng J Y Li C J 2018 Metrologia 55 360
[11] Huang P W Tang B Chen X Zhong J Q Xiong Z Y Zhou L Wang J Zhan M S 2019 Metrologia 56 045012
[12] Snadden M J McGuirk J M Bouyer P Haritos K G Kasevich M A 1998 Phys. Rev. Lett. 81 971
[13] Duan X C Zhou M K Mao D K Yao H B Deng X B Luo J Hu Z K 2014 Phys. Rev. 90 023617
[14] Wang Y P Zhong J Q Song H W Zhu L Li Y M Chen X Li R B Wang J Zhan M S 2017 Phys. Rev. 95 053612
[15] Canuel B Leduc F Holleville D Gauguet A Fils J Virdis A Clairon A Dimarcq N Borde C J Landragin A Bouyer P 2006 Phys. Rev. Lett. 97 010402
[16] Tackmann G Berg P Schubert C Abend S Gilowski M Ertmer W Rasel E M 2012 New J. Phys. 14 015002
[17] Dickerson S M Hogan J M Sugarbaker A Johnson D M S Kasevich M A 2013 Phys. Rev. Lett. 111 083001
[18] Yao Z W Lu S B Li R B Wang K Cao L Wang J Zhan M S 2016 Chin. Phys. Lett. 33 083701
[19] Yao Z W Lu S B Li R B Luo J Wang J Zhan M S 2018 Phys. Rev. 97 013620
[20] Fray S Diez C A Hänsch T W Weitz M 2004 Phys. Rev. Lett. 93 240404
[21] Bonnin A Zahzam N Bidel Y Bresson A 2013 Phys. Rev. 88 043615
[22] Schlippert D Hartwig J Albers H Richardson L L Schubert C Roura A Schleich W P Ertmer W Rasel E M 2014 Phys. Rev. Lett. 112 203002
[23] Tarallo M G Mazzoni T Poli N Sutyrin D V Zhang X Tino G M 2014 Phys. Rev. Lett. 113 023005
[24] Zhou L Long S T Tang B Chen X Gao F Peng W C Duan W T Zhong J Q Xiong Z Y Wang J Zhang Y Z Zhan M S 2015 Phys. Rev. Lett. 115 013004
[25] Barrett B Antoni-Micollier L Chichet L Battelier B Lévèque T Landragin A Bouyer P 2016 Nat. Commun. 7 13786
[26] Duan X C Zhou M K Deng X B Yao H B Shao C G Luo J Hu Z K 2016 Phys. Rev. Lett. 117 023001
[27] Rosi G D’Amico G Cacciapuoti L Sorrentino F Prevedelli M Zych M Brukner Č Tino G M 2017 Nat. Commun. 8 15529
[28] Zhang K Zhou M K Cheng Y Chen L L Luo Q Xu W J Cao L S Duan X C Hu Z K 2020 Chin. Phys. Lett. 37 043701
[29] Zhou L He C Yan S T Chen X Duan W T Xu R D Zhou C Ji Y H Bathwal S Wang Q Hou Z Xiong Z Y Gao D F Zhang Y Z Ni W T Wang J Zhan M S 2019 arXiv:1904.07096 [quant-ph]
[30] Freier C Hauth M Schkolnik V Leykauf B Schilling M Wziontek H Scherneck H G Müller J Peters A 2016 J. Phys.: Conf. Ser. 723 012050
[31] Fu Z J Wu B Cheng B Zhou Y Weng K Zhu D Wang Z Lin Q 2019 Metrologia 56 025001
[32] Wu X Pagel Z Malek B S Nguyen T H Zi F Scheirer D S Müller H 2019 Sci. Adv. 5 eaax0800
[33] Peters A Chung K Y Chu S 2001 Metrologia 38 25
[34] Louchet-Chauvet A Farah T Bodart Q Clairon A Landragin A Merlet S Pereira Dos Santos F 2011 New J. Phys. 13 065025
[35] Lan S Y Kuan P C Estey B Haslinger P Müller H 2012 Phys. Rev. Lett. 108 090402
[36] Bongs K Launay R Kasevich M A 2006 Appl. Phys. 84 599
[37] Farah T Gillot P Cheng B Landragin A Merlet S Pereira Dos Santos F 2014 Phys. Rev. 90 023606
[38] Hogan J M Johnson D M S Kasevich M A 2008 arXiv:0806.3261v1 [physics.atom-ph]
[39] Zhou L Xiong Z Y Yang W Tang B Peng W C Hao K Li R B Liu M Wang J Zhan M S 2011 Gen. Relat. Gravit. 43 1931
[40] Foster G T Fixler J B McGuirk J M Kasevich M A 2002 Opt. Lett. 27 951